Model Number 5392
Device Problem Display or Visual Feedback Problem (1184)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/22/2020
Event Type  malfunction  
Manufacturer Narrative
Product analysis : it was noted on analysis that the liquid crystal display (lcd) backlight would not turn on as there was an issue identified with the printed circuit board (pcb) connector.It was also noted that the ventricular output connector was contaminated inside and outside of connector.The keypad key dome was collapsing and the keypad window was damaged.Encoder assembly and three control knobs were contaminated with rust.The lower case was contaminated and broken.The main seal, all six case screws and hanger screws were contaminated.Both output connector nuts were discolored.Both wires on the lcd were pinched but not compromised.Device errors were identified.Firmware was updated and the pcb was reset.All found defective parts were replaced and all other identified issues were resolved.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that an external pulse generator (epg) experienced an error.The epg was returned for service.No patient complications have been reported as a result of this event.It was further reported that the programmer subsequently tested out of specification during manufacturer's analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUAL CHAMBER TEMPORARY PACEMAKER
Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10390038
MDR Text Key202382565
Report Number3004593495-2020-00843
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2020
Date Manufacturer Received07/20/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-