Product analysis : it was noted on analysis that the liquid crystal display (lcd) backlight would not turn on as there was an issue identified with the printed circuit board (pcb) connector.It was also noted that the ventricular output connector was contaminated inside and outside of connector.The keypad key dome was collapsing and the keypad window was damaged.Encoder assembly and three control knobs were contaminated with rust.The lower case was contaminated and broken.The main seal, all six case screws and hanger screws were contaminated.Both output connector nuts were discolored.Both wires on the lcd were pinched but not compromised.Device errors were identified.Firmware was updated and the pcb was reset.All found defective parts were replaced and all other identified issues were resolved.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
|