• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SET MENISCUS MENDER II DISPOSABLE; PASSER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SET MENISCUS MENDER II DISPOSABLE; PASSER Back to Search Results
Model Number 7209485
Device Problem Material Separation (1562)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/21/2020
Event Type  malfunction  
Event Description
It was reported that during the meniscus repair surgery, the loop retriever was already disassembled between head and shaft when the package was opened.A backup device was available to complete the procedure with no delay or patient injuries.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3, h6: one 7209485 disposable meniscus mender ii set was returned for evaluation.The visual assessment confirmed the complaint.The loop meniscal were broken from the bottom.To avoid components from shifting within the package, individual component storage cradles are intentionally snug.Use of the distal end (head) to remove snare loops from the tray may result in weakening, bending or complete fracture between at the head and shaft connection.Proper method of retrieval is to push and pop the head of the component from the back of the cradle.Complaint history review was unattainable without a valid lot number and product code provided although query using entire mender family indicated similar allegations.Batc review was unattainable without a valid lot number.Engineering evaluation confirmed the product met specifications at the time of distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SET MENISCUS MENDER II DISPOSABLE
Type of Device
PASSER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key10391173
MDR Text Key202395227
Report Number1219602-2020-01227
Device Sequence Number1
Product Code HWQ
UDI-Device Identifier03596010468482
UDI-Public03596010468482
Combination Product (y/n)N
PMA/PMN Number
K885311
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number7209485
Device Catalogue Number7209485
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2020
Date Manufacturer Received10/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-