Model Number OER-PRO
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Further investigation from the field service engineer identified the connector was not broken but loose.The unit was repaired (resealed) in the field and verified according to oem instructions.
 
Event Description
It was reported the gray connector of the reprocessor was broken.
 
Manufacturer Narrative
This report is being supplemented to provide corrected data regarding the reported event.After further review it was determined the representative identified the loose connection on (b)(6) 2020.Therefore, the aware date of this event is (b)(6) 2020 and not (b)(6)2020 as previously reported.
 
Manufacturer Narrative
The investigation has been completed.A device history record (dhr) was performed and found no non-conformances that would cause the reported malfunction.All records indicate that the device was manufactured in accordance with all applicable procedures.Instructions for use (ifu) states: ¿do not use the equipment if any connector seems to be damaged or defective.Using the equipment when an irregularity has been detected may interfere with reprocessing.Furthermore, fluid leakage may damage peripheral device or facilities near the equipment.¿ the cause of the event cannot be conclusively determined.Possible causes include: an impact that was caused to the unit causing it to become loose or to come apart.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10398721
MDR Text Key202686179
Report Number8010047-2020-05457
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup,Followup
Report Date 10/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-