Model Number 8637-20
Device Problems Premature End-of-Life Indicator (1480); Premature Elective Replacement Indicator (1483)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/24/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a foreign consumer via a manufacturer representative regarding a patient receiving intrathecal baclofen 1000 mcg/ml at 140. 2 mcg/day via an implantable pump. It was reported that the pump was refilled on (b)(6) 2020, where it still had 17 months left until elective replacement indicator (eri). On a later date, the patient heard a non-critical alarm. Logs showed that premature eri occurred on (b)(6) 2020. No motor stalls or low battery resets had occurred. The pump was replaced. The customer discarded the pump. The issue was resolved. The patient's status was alive - no injury. It was noted that the pump was implanted in (b)(6) 2015 (no specific date reported). There were no environmental, external or patient factors that might have led or contributed to the event. Information regarding the patient¿s gender, age, weight, medical history, and other medications was unavailable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNCHROMED II
Type of DevicePUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MDT PUERTO RICO OPERATIONS CO
rd 31 km 24 hm 4
juncos PR 00777
Manufacturer (Section G)
MDT PUERTO RICO OPERATIONS CO
rd 31 km 24 hm 4
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key10398941
MDR Text Key202696035
Report Number3004209178-2020-13970
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 08/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/12/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date05/14/2015
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/17/2020
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured11/22/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 08/12/2020 Patient Sequence Number: 1
-
-