A physician reported csf retention and a "skin bump¿: the certas valve was implanted via v-p shunt by end of (b)(6) 2020 with an unknown setting.However, within a month from the procedure, a skin bump was observed at valve implantation site, csf retention was found, and valve detachment was suspected.Therefore, the valve was replaced to a new one on (b)(6) 2020.
|
Updated fields: d10, g4, g7, h2, h3, h6, h10.Udi - (b)(4).The valve was returned for evaluation: review of the history device records was not possible as the lot number was unknown.Failure analysis - the valve was visually inspected; the silicone housing was cut/torn around the siphon guard.The valve passed the test for programming, occlusion, siphon guard and pressure.The valve was leak tested; leaked from the torn/cut silicone housing.The catheters were irrigated no occlusions noted.The root cause for the cut/tear in the silicone housing was probably due to a sharp or pointed object coming into contact with the silicone housing, as noted in the ifu silicone has a low cut/tear resistance.The possible root cause for the issue reported by the customer was due to the cut/torn silicone housing, csf fluid was not flowing due to the cut/torn silicone housing hence causing a swelling (bump).
|