LIVANOVA DEUTSCHLAND GMBH HEATER-COOLER SYSTEM 3T CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problem Defective Component (2292)
Patient Problem No Patient Involvement (2645)
Event Date 07/17/2020
Event Type  Malfunction  
Manufacturer Narrative

Was no known patient involvement. Livanova deutschland manufactures the heater-cooler system 3t. The incident occurred in (b)(6). Field service engineer replaced distribution board and stirrer motor and then tested the device extensively with all test results ok. If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.

 
Event Description

Livanova received a report about an heater cooler with replacement of distribution board and stirrer motor. There is no evidence of patient involvement.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM 80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key10408019
MDR Text Key204026019
Report Number9611109-2020-00467
Device Sequence Number1
Product Code DWC
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL
Reporter Occupation
Type of Report Initial,Followup
Report Date 07/17/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received08/14/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator NO INFORMATION
Device MODEL Number16-02-85
Was Device Available For Evaluation? No Answer Provided
Is The Reporter A Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received09/19/2020
Was Device Evaluated By Manufacturer? No Answer Provided
Is The Device Single Use? No Answer Provided
Is this a Reprocessed and Reused Single-Use Device?
Type of Device Usage Invalid Data

-
-