• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LENS 4K A/C CAMERA HEAD; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LENS 4K A/C CAMERA HEAD; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 72205058
Device Problem Erratic or Intermittent Display (1182)
Patient Problem No Patient Involvement (2645)
Event Date 07/23/2020
Event Type  malfunction  
Event Description
It was reported that the camera head was plugged into the camera control unit and the image on the monitor keeps flickering.The incident occurred before the procedure; no delay, backup device available.No complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10, d4, h2, h3, h6: the reported device, intended for use in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.¿a complaint history review concluded this was a repeat issue.A customer provided image confirms the product information, but does not provide functional information.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LENS 4K A/C CAMERA HEAD
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10410541
MDR Text Key203155584
Report Number1643264-2020-00855
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier00885556720790
UDI-Public00885556720790
Combination Product (y/n)N
PMA/PMN Number
K191177
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup
Report Date 09/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72205058
Device Catalogue Number72205058
Was Device Available for Evaluation? No
Date Manufacturer Received09/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-