Device evaluation of battery charger sn (b)(4) has been completed.The reported problem (battery charger problem) has been confirmed.Upon investigation the battery charger was unable to fully power on.The charger exhibited a flash memory failure at bga components u102 and u105 on the c/a board.Additionally, contamination was found on the battery and table boards.The root cause for the flash memory failure could not be positively identified.The root cause for the contamination was ingress of an unknown liquid.Lifevest patient training materials have been updated as a reminder not to expose the lifevest electronic components to liquids.No adverse event resulted from the defective battery charger.
|