• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. CARELINK PROGRAMMER; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. CARELINK PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090
Device Problems Overheating of Device (1437); Use of Device Problem (1670); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/08/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmer had a slight burning smell.The programmer was powered off for several minutes because it was thought it might have overheated.When an attempt was made to restart the programmer a faint crackling or hissing noise was heard.The current status of the programmer is unknown.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product analysis: the analysis could not confirm the customer comment of the programmer had a slight burning smell.The programmer booted up without issue.The programmer was disassembled no anomalies, no problems were observed or found.The programmer works normally.It was also determined at analysis that the latch of the cable bay is broken.The cable of the radiofrequency head is worn however functions correctly, will be replaced as a preventative measure.The bullets assy is missing and the anti-slip layer is ripped off the label of the radiofrequency head.The bezel has minor damage, considered as acceptable.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
MDR Report Key10419536
MDR Text Key203288255
Report Number2182208-2020-01546
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 10/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2020
Date Manufacturer Received09/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
2067L RF HEAD/CABLE-14 FT; 2067L RF HEAD/CABLE-14 FT
-
-