WILLIAM COOK EUROPE COOK CELECT FEMORAL AND JUGULAR VENA CAVA FILTER; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number IGTCFS-65-UNI-CELECT-PERM
Device Problems Structural Problem (2506); Device Tipped Over (2589); Unintended Movement (3026); Insufficient Information (3190)
Patient Problems Chest Pain (1776); Dyspnea (1816); Perforation of Vessels (2135); No Information (3190)
Event Type  Injury  
Manufacturer Narrative
Manufacturer ref# (b)(4).Blank fields on this form indicate the information is unknown or unavailable.Catalog# is unknown but referred to as cook celect filter.Occupation: non-healthcare professional.It has not been possible to investigate or evaluate this alleged event based on the limited information provided to date.Cook will reopen its investigation if further information is receiving warranting supplementation in accordance with 21 c.F.R.803.56.
 
Event Description
Description of event according to short form complaint filed: it is alleged that "[pt] received a cook celect filter on (b)(6) 2013.It is alleged that the [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.".
 
Manufacturer Narrative
The following fields were updated per additional information received: a4, b1, b2, b5, b6, b7, d1, d4, g5, and h6.Investigation: the following allegations have been investigated: vena cava perforation, migration, tilt, chest pain, shortness of breath.The reported allegations have been investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter or filter fragment migration and (or) embolization (e.G., movement to the heart or lungs) has been reported.Filter or filter fragment movement has occurred in both the cranial and caudal direction and may be either symptomatic or asymptomatic.Potential causes may include filter placement in ivcs with diameters smaller or larger than those specified in these instructions for use; improper deployment; deployment into thrombus; dislodgement due to large thrombus burdens; and (or) excessive force or manipulations near an in situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: filter migration, trauma to adjacent structures.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported chest pain and shortness of breath are directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog number and lot number are unknown.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
Patient allegedly received an implant on (b)(6) 2013 via right common femoral vein due to post deep vein thrombosis.The patient alleges migration, tilt, and vena cava perforation.The patient further alleges chest pain and shortness of breath.Per certificate of death, dated (b)(6) 2020, immediate cause of death is septic shock due to urinary tract infection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK CELECT FEMORAL AND JUGULAR VENA CAVA FILTER
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key10423905
MDR Text Key203461641
Report Number3002808486-2020-00782
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
PMA/PMN Number
K073374
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 10/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/19/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberIGTCFS-65-UNI-CELECT-PERM
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received09/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Weight91
-
-