COVIDIEN MFG SOLUTIONS S.A. MAHURKAR; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
|
Back to Search Results |
|
Model Number 8813793009 |
Device Problem
Product Quality Problem (1506)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/03/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during use, the butterfly clasp (tip of the catheter) was not tight.The catheter was not repaired.There was no tego utilized and luer adapter issue.It was stated that there was no blood loss.A new catheter was used to resolve the issue.There was no patient injury.
|
|
Event Description
|
According to the reporter, during use, the butterfly clasp (tip of the catheter) was not tight.A new catheter was used to complete the treatment.There was no stitches.There was no unnecessary movements contributed to the tightening of the butterfly clasp.The catheter was treated with prior to product placement.No other intervention or treatment required.The catheter was not repaired.There was no tego utilized and luer adapter issue.There was no blood loss and no blood transfusion required.There was no packaging damage, nothing unusual observed prior to opening the package, no other defects or damage found on the product there was no patient injury.
|
|
Manufacturer Narrative
|
Additional information: b5, g4.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during use, the butterfly clasp (tip of the catheter) was not tight.A new catheter was used to complete the treatment.It was stated that there was no stitches.There was no unnecessary movements contributed to the tightening of the butterfly clasp.The catheter was treated with prior to product placement.No other intervention or treatment required.The catheter was not repaired.There was no tego utilized and luer adapter issue.There was no blood loss and no blood transfusion required.There was no packaging damage, nothing unusual observed prior to opening the package, no other defects or damage found on the product there was no patient injury.
|
|
Manufacturer Narrative
|
H3: evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the catheter was blood stained with dried blood inside the distal end of the cannula.The suture wing was not received.The red and blue luer adapters, extension tubes, clamps, hub and cannula appeared intact.Functional testing found no loose components.A test guide wire was inserted into both lumen, it passed through and was removed without restriction.The catheter was submerged into a water bath.The ends were clamped, and a syringe was used to inject air to observe leakage.No air bubbles were present.Both extensions were tested with acceptable results.It was reported that there was a securement wing issue.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|