ADVANCED BIONICS, LLC HIRES ULTRA 3D IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1601-05
Device Problem Migration (4003)
Patient Problems Hearing Impairment (1881); Failure of Implant (1924); Therapeutic Effects, Unexpected (2099); Dizziness (2194)
Event Date 07/23/2020
Event Type  Injury  
Event Description
The recipient reportedly experienced device migration.The recipient presented with sound quality issues, non-auditory sensations, and dizziness.A ct scan confirmed device migration.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed silicone damage on the top cover, a damaged electrode ground ring, and the electrode was severed near the array prior to receipt.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES ULTRA 3D IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key10426355
MDR Text Key203528128
Report Number3006556115-2020-00818
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016878008
UDI-Public(01)07630016878008(11)200228(17)230228
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup,Followup
Report Date 07/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/19/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Expiration Date02/28/2023
Device Model NumberCI-1601-05
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age60 YR
-
-