Model Number N/A
Device Problem Component Missing (2306)
Patient Problem No Patient Involvement (2645)
Event Date 07/29/2020
Event Type  malfunction  
Manufacturer Narrative
The manufacturer has received the sample and is pending evaluation.Results are expected soon.A lot history review (lhr) of asdss0184 showed no other similar product complaint(s) from this lot number.
 
Event Description
It was reported that the needle protection cap was missing.There was no reported patient involvement.
 
Manufacturer Narrative
H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a missing needle cover was inconclusive due to the condition of the returned sample.The product returned for evaluation was one 22ga x 0.75¿ safestep safety infusion set.The sample was received with its original opened packaging.The clear plastic needle cover was not present on the needle or in the packaging.While no needle cover was observed, the opened state of the packaging prevented determination of the original state of the device.Consequently this complaint is inconclusive at this time.A lot history review (lhr) of asdss0184 showed no other similar product complaint(s) from this lot number.
 
Event Description
It was reported that the needle protection cap was missing.There was no reported patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAFESTEP HUBER NEEDLE SET 22G X 0.75IN WITH Y-SITE
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key10434615
MDR Text Key204370154
Report Number3006260740-2020-02992
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00801741066177
UDI-Public(01)00801741066177
Combination Product (y/n)N
PMA/PMN Number
K153440
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 09/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/20/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/28/2022
Device Model NumberN/A
Device Catalogue NumberLH-0029YN
Device Lot NumberASDSS0184
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/17/2020
Event Location Hospital
Date Manufacturer Received08/25/2020
Is This a Reprocessed and Reused Single-Use Device? No