RESPIRONICS CALIFORNIA, INC V60 VENTILATOR VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problems
Blower (738); Insufficient Cooling (1130)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
Malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.
Date of report: 20aug2020.
|
|
Event Description
|
It was reported to philips by the customer (biomed) that the v60 ventilator displayed an over temperature error.
The device was not being used on a patient at the time of the event.
There was no adverse event to the patient or user as a result of this event.
The device was evaluated by the customer with assistance from a philips remote service engineer (rse).
The rse advised the customer to verify the v60 cooling fan is running, and to verify the air inlet filter is clean.
The rse also asked the biomed to check the significant event log for over temperature errors, however the biomed is unable to locate any temperature errors in the log.
A philips field service engineer (fse) was dispatched to the customer site.
The reported issue was confirmed and traced to a faulty blower.
The blower assembly was replaced to resolve the reported issue.
|
|