• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. CARELINK PROGRAMMER; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. CARELINK PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090
Device Problems Display or Visual Feedback Problem (1184); Mechanical Problem (1384); Device Difficult to Program or Calibrate (1496)
Patient Problem No Patient Involvement (2645)
Event Date 08/07/2020
Event Type  malfunction  
Manufacturer Narrative
Product analysis: analysis was able to confirm the customer comment that programmers touch pen pointer was displaced about 1 cm below the screen and that the handle support was cracked.After cleaning the xy board, performed 25 point calibration.The handle support was replaced.All found defective parts were replaced and all other identified issues were resolved.The device then passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmers touch pen pointer was displaced about 1 centimetre below the screen.It was further noted that the handle support was cracked.The device has been returned to service.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10435827
MDR Text Key203799947
Report Number2182208-2020-01583
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/21/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2020
Date Manufacturer Received08/07/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/28/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
2076 RADIOFREQUENCY HEAD.
-
-