SMITH & NEPHEW, INC. JRNY II TKA FEM TRIAL LT SZ 5; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/META
|
Back to Search Results |
|
Model Number 74031225 |
Device Problem
Nonstandard Device (1420)
|
Patient Problems
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/29/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during a training lab at the office, it was noted that there is a burr in the sliding track of the trial.The box cut guide will not properly slide into place because of this burr.Also, the device exhibits signs of wear and tear.No surgery or patient were involved.
|
|
Manufacturer Narrative
|
The device, intended use in treatment, was returned for evaluation.A visual inspection confirmed the rails are cracked on the jrny ii tka fem trial lt sz 5.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A review of complaint history for the listed part revealed no prior complaints for the listed batch mode.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
|
|
Search Alerts/Recalls
|
|
|