Model Number 72200616
Device Problem Electrical Shorting (2926)
Patient Problem No Patient Involvement (2645)
Event Date 06/15/2020
Event Type  Malfunction  
Manufacturer Narrative

There was no relationship found between the returned device and the reported incident. A visual inspection was performed on the product and observed no issues. A functional evaluation revealed a jammed motor. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended. A complaint history review concluded this was a repeat issue. The complaint was not verified and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process. Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.

 
Event Description

It was reported that the motor drive unit had a short circuit. No case reported; therefore, there was no patient involvement. Results of investigation have concluded that this unit presented a jammed motor which makes it a reportable event. All available information has been disclosed. If additional information should become available, a supplemental report will be submitted accordingly.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMOTOR DRIVE UNIT HAND CNTRL PWRMX EL
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key10439479
MDR Text Key204016912
Report Number1643264-2020-00924
Device Sequence Number1
Product Code HAB
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberEXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,HEALTH
Reporter Occupation OTHER HEALTH CARE PROFESSIONAL
Type of Report Initial
Report Date 08/21/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received08/21/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number72200616
Device Catalogue Number72200616
Was Device Available For Evaluation? Device Returned To Manufacturer
Date Returned to Manufacturer06/24/2020
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/07/2020
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-