Model Number OER-2
Device Problem Device Reprocessing Problem (1091)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/28/2020
Event Type  malfunction  
Manufacturer Narrative
Since the subject device has not been returned to olympus medical systems corp.(omsc) for the evaluation, the exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that the user facility had not reprocessed endoscopes after use each time but had reprocessed endoscopes at the end of the day with the subject device.The user facility would not like to take around 20 minutes to reprocess and keep to use the endoscopes without the reprocessing.Olympus representative will train the facility for the appropriate reprocessing with the subject device in the near future.There was no report of patient injury associated with this event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.Since the subject device was not returned to olympus medical systems corp.(omsc), it could not be investigated.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.However, omsc surmised there was the possibility of the inappropriate or insufficient cleaning and disinfection for the subject device.If additional information becomes available, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OER-2 100V
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10440573
MDR Text Key223646683
Report Number8010047-2020-05731
Device Sequence Number1
Product Code FEB
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/14/2020
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/28/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-