Model Number DACAPO POWERPACK |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problems
No Information (3190); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been returned to innsbruck where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
|
Event Description
|
A broken and leaking dacapo powerpack has been received at service repair.
|
|
Manufacturer Narrative
|
Conclusion: the returned device was visually inspected upon arrival.A mechanically damaged housing was observed.The observed damage did not allow electrical testing to be conducted.The damage to the battery housing was clearly the reason for the malfunction of the device returned by the end-user.This is a final report.
|
|
Event Description
|
A broken and leaking dacapo powerpack has been received at service repair.
|
|