• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM; EXTERNAL COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM; EXTERNAL COMPONENT Back to Search Results
Model Number DACAPO POWERPACK
Device Problem Fluid/Blood Leak (1250)
Patient Problems No Information (3190); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Type  malfunction  
Manufacturer Narrative
The device has been returned to innsbruck where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
 
Event Description
A broken and leaking dacapo powerpack has been received at service repair.
 
Manufacturer Narrative
Conclusion: the returned device was visually inspected upon arrival.A mechanically damaged housing was observed.The observed damage did not allow electrical testing to be conducted.The damage to the battery housing was clearly the reason for the malfunction of the device returned by the end-user.This is a final report.
 
Event Description
A broken and leaking dacapo powerpack has been received at service repair.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Type of Device
EXTERNAL COMPONENT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
MDR Report Key10441197
MDR Text Key204007245
Report Number9710014-2020-00461
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P000025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 09/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDACAPO POWERPACK
Device Lot NumberSN: 1006D044
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2020
Was the Report Sent to FDA? No
Date Manufacturer Received08/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age22 YR
-
-