Catalog Number 5111-00200-060
Device Problem Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/02/2020
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a 2.0mm coupler device had a steel needle on one side of the anastomotic ring that was bent.This issue was identified when opening the package for use.A new coupler was used to complete procedure.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information was added to d10, h3, and h6.H10: the device was received for evaluation.A visual inspection was performed, and it was noted that the rings showed signs of use (dried blood) which is consistent with the complaint¿s statement that the bent pin was found during use.When viewing the right ring of the 2.0mm coupler jaw assembly, a bent pin was visible.The reported condition was verified.The cause of the reported condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.0MM COUPLER
Type of Device
DEVICE, ANASTOMOTIC, MICROVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10445775
MDR Text Key204164501
Report Number1416980-2020-05159
Device Sequence Number1
Product Code MVR
UDI-Device Identifier10844735002642
UDI-Public10844735002642
Combination Product (y/n)Y
PMA/PMN Number
K861985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number5111-00200-060
Device Lot NumberSP19G01-1382542
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2020
Date Manufacturer Received09/17/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-