MEDTRONIC EUROPE SARL PERCEPTA CRT-P MRI SURESCAN; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
|
Back to Search Results |
|
Model Number W1TR01 |
Device Problem
Low impedance (2285)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/05/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: 1388t x2, leads, implanted: (b)(6) 1999.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that there was a low unipolar impedance warning on the right ventricular (rv) lead when the device did not actually measure a low impedance.The cardiac resynchronization therapy defibrillator (crt-d) remains in use.There was also possible high threshold on the left ventricular (lv) lead.The lv lead remains in use.No patient complications have been reported as a result of this event.
|
|
Event Description
|
It was reported that there was a low unipolar impedance warning on the right ventricular (rv) lead when the device did not actually measure a low impedance.The cardiac resynchronization therapy pacemaker (crt-p) remains in use.There was also possible high threshold on the left ventricular (lv) lead.The lv lead remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated a false alert for lead impedance out of range.Correction: b5.Desc evt problem medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received indicated there was another noted alert for low unipolar impedance warning on the right ventricular (rv) lead and an alert for low unipolar impedance warning on the right atrial (ra) lead when the device did not actually measure a low impedance.It was also noted that the lv lead exhibited high threshold.
|
|
Manufacturer Narrative
|
Product event summary the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated a false alert for lead impedance out of range.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|