• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. FIBEROPTIX ULTRA 8 IAB: 8FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. FIBEROPTIX ULTRA 8 IAB: 8FR 40CC; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN000254
Device Problem Material Puncture/Hole (1504)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/08/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported the intra-aortic balloon (iab) was inserted into the patient and after 10 minutes it was noted that the balloon ruptured in the patient.As a result, the iab and sheath was removed and a new iab and sheath was used, using the same insertion site.There was no report of patient complications, serious injury or death.
 
Manufacturer Narrative
(b)(4) teleflex received the device for investigation.The reported complaint of iab leak suspected is confirmed.A puncture to the bladder, consistent with contact from the broken fiber, was found near the distal tip of the catheter which allowed blood to enter the helium pathway.The root cause of the broken fiber is undetermined.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revived risk.This will be monitored for any developing trends.
 
Event Description
It was reported the intra-aortic balloon (iab) was inserted into the patient and after 10 minutes it was noted that the balloon ruptured in the patient.As a result, the iab and sheath was removed and a new iab and sheath was used, using the same insertion site.There was no report of patient complications, serious injury or death.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBEROPTIX ULTRA 8 IAB: 8FR 40CC
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key10453602
MDR Text Key204359717
Report Number3010532612-2020-00238
Device Sequence Number1
Product Code DSP
UDI-Device Identifier00801902007247
UDI-Public00801902007247
Combination Product (y/n)N
PMA/PMN Number
K021462
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/26/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2022
Device Model NumberIPN000254
Device Catalogue NumberIAB-05840-LWS
Device Lot Number18F20C0029
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/26/2020
Date Manufacturer Received09/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-