STRYKER ENDOSCOPY-SAN JOSE PKG., AGG. PLUS SHAVER BLADE, F- SERIES - 4.0MM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
|
Back to Search Results |
|
Catalog Number 0375544000 |
Device Problems
Break (1069); Fracture (1260)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/28/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional information will be provided once the investigation has been completed.
|
|
Event Description
|
It was reported that the tip broke during the procedure.The piece was retrieved.
|
|
Manufacturer Narrative
|
Alleged failure: shaver break off in a patient today (tip).The failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be the blade comes in contact with a hard object causing damage to the teeth and hitting the housing edges.Excessive pressure, such as bending or prying, may cause the arthroscopic shaver blades to bend/break.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
|
Event Description
|
It was reported that the tip broke during the procedure.The piece was retrieved.
|
|