ZOLL CIRCULATION AUTOPULSE PLATFORM RESUSCITATION; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Patient Involvement (2645)
Event Date 08/03/2020
Event Type  malfunction  
Manufacturer Narrative
The reported complaint of autopulse platform (sn (b)(4)) displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) was confirmed during initial functional testing and archive data review.The root cause for ua 07 was due to defective load cells.The defective load cells were likely due to a defective component or mishandling such as drop.During visual inspection, no physical damage was observed.During archive data review, multiple ua 07 error messages were recorded around the reported event date, thus confirming the reported complaint.During the functional testing, the platform's load sensing system has detected a weight/load imbalance between the two load cells (checked during power-on-self-test).Load cell characterization results confirmed load cells over reported (defective).Thus, confirming the reported complaint.To remedy ua 07, damaged load cells were replaced.After replacing load cells, the autopulse was subjected to the run-in test using the 95% patient large resuscitation test fixture (lrtf) with good known test battery until discharged without any fault or error.The load cell characterization test passed.The autopulse platform passed final test.Historical complaints were reviewed for service information related to the reported complaint and there was no similar complaint reported for autopulse with serial number (b)(4).
 
Event Description
During device check, the autopulse platform (sn (b)(4)) displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) error message.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE PLATFORM RESUSCITATION
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose, ca
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave,
san jose, ca 
4192922
MDR Report Key10463761
MDR Text Key205013429
Report Number3010617000-2020-00826
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/27/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/04/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/22/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-