• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 GAS BLENDER SYSTEM; GAS CONTROL UNIT, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 25-40-45
Device Problem Gas Output Problem (1266)
Patient Problem Low Oxygen Saturation (2477)
Event Date 07/22/2020
Event Type  Injury  
Manufacturer Narrative
Patient information was not provided.Livanova deutschland manufactures the s5 gas blender system.The incident occurred in (b)(6).The device was requested back to the manufacturer site for further investigation.During functional tests deviation of the co2 output and a/d converter fault on air display could be identified.Further testing confirmed the same findings and one more error code associated to a discrepancy between the set and the actual value of air was experienced during pre-calibration phase.The co2 mass flow controller, air mass flow controller, the honeywell sensor and cables harness for air/o2/co2 will be replaced and a re-calibration will be performed.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that the blood of the patient became darker during procedure and the s5 gas blender system was changed.The procedure could be completed successfully with no further impact to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 GAS BLENDER SYSTEM
Type of Device
GAS CONTROL UNIT, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key10464263
MDR Text Key204698739
Report Number9611109-2020-00482
Device Sequence Number1
Product Code DTX
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K101046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 08/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/28/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number25-40-45
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/29/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/22/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-