Patient information was not provided.Livanova deutschland manufactures the s5 gas blender system.The incident occurred in (b)(6).The device was requested back to the manufacturer site for further investigation.During functional tests deviation of the co2 output and a/d converter fault on air display could be identified.Further testing confirmed the same findings and one more error code associated to a discrepancy between the set and the actual value of air was experienced during pre-calibration phase.The co2 mass flow controller, air mass flow controller, the honeywell sensor and cables harness for air/o2/co2 will be replaced and a re-calibration will be performed.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|