It was reported that the normothermia patient was on arctic sun device.
They were getting a low flow alert (alert 02).
The pads may be on due to be changed.
The nurse attempted to empty the pads, but received alert 07 (empty pads cycle not complete).
Disconnected pads and placed device in manual mode.
Fluid delivery line showed flow rate of 1.
7lpm, inlet pressure of 7.
0psi, circulation pump command 45%.
Left chest flow rate was 1.
6lpm, inlet pressure was 7.
1psi, circulation pump command 45%.
Right chest flow rate of 1.
7lpm, inlet pressure -7.
1psi, circulation pump command 45%.
Left thigh flow rate of 2.
2lpm, inlet pressure -7.
3psi, circulation pump command 52%.
Right thigh: flow rate 2.
1lpm, inlet pressure -7.
0psi, circulation pump command 59%.
System hours 3794.
Pump hours 3301.
Placed device back on automatic mode.
Flow 2.
1lpm.
While testing the flow of the device it gave a low patient temperature alert.
Patient temperature read 15.
3c.
The foley probe was on use.
Nurse adjusted the probe and now temperature reads 36.
2c.
Explained that the probe or temperature in cable needs to be changed should the patient's temperature drop like that again.
Nurse stated that the device accidentally came unplugged from the wall earlier and thought that was when the alerts began.
|