Product analysis: analysis of the external pulse generator (epg) was able to confirm the customer comments of broken output connectors.Analysis also noted that the hanger was broken, backlight issue due to loose connector on main printed circuit board (pcb), lower case boss was broken, and 4 case screws were contaminated.All found defective parts were replaced and all other identified issues were resolved.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
|