Device Problem Break (1069)
Patient Problem Perforation (2001)
Event Date 08/28/2019
Event Type  Injury  
Event Description

It was reported that a perforation occurred. A greenfield filter was implanted on (b)(6) 1996. On (b)(6) 2019 the patient had a ct scan of their chest and abdomen. The imaging showed that multiple struts had perforated the inferior vena cava (ivc), with three of those struts fracturing, and one embedded in the patients vertebrae. On (b)(6) 2019, the patient underwent a failed standard percutaneous retrieval attempt. The fractured greenfield filter was retrieved, but two of the fractured struts remained embedded within and extending beyond the ivc and the third strut unaccounted for.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameGREENFIELD
Type of DeviceFILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key10471029
MDR Text Key204878033
Report Number2134265-2020-11735
Device Sequence Number1
Product Code DTK
Combination Product (Y/N)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type CONSUMER
Reporter Occupation
Type of Report Initial
Report Date 08/31/2020
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received08/31/2020
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? No
Device Operator HEALTH PROFESSIONAL
Was Device Available For Evaluation? No
Is The Reporter A Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/11/2020
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device?
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 08/31/2020 Patient Sequence Number: 1
-
-