MAUDE Adverse Event Report: ABBOTT VASCULAR XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR XIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Catalog Number 1070350-12
Device Problems Improper or Incorrect Procedure or Method (2017); Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Patient Problem Vessel Or Plaque, Device Embedded In (1204)
Event Date 08/07/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4). The stent remains in the patient. Investigation is not yet complete. A follow up report will be submitted with all additional relevant information.
 
Event Description
It was reported the procedure was to treat a moderately tortuous, and moderately calcified in stent restenosis (80%) in the distal left circumflex artery (lcx). The 3. 50x12mm xience xpedition stent delivery system (sds) was advanced; however failed to cross the 40% stenosed proximal part of the lcx due to resistance with the anatomy. A buddy wire was used to attempt to re-advance the sds but this also failed. After sometime of maneuvering it was noted in the angiogram that the stent dislodged from the sds before deployment and was hanging in the proximal lcx. A 3. 5x13mm non-abbott stent was advanced and deployed at high pressure apposing the stent to the vessel wall. Timi iii flow was achieved. There was no reported clinically significant delay in the procedure and the patient was reported to be stable. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameXIENCE XPEDITION EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Type of DeviceDRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key10472773
MDR Text Key204988605
Report Number2024168-2020-07232
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeIN
PMA/PMN Number
P110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Type of Report Initial,Followup
Report Date 09/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date06/06/2022
Device Catalogue Number1070350-12
Device Lot Number9052941
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/09/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/07/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 08/31/2020 Patient Sequence Number: 1
-
-