STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR FEM COMP #3 L-CEM PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5510-F-301
Device Problem Loss of Osseointegration (2408)
Patient Problem Injury (2348)
Event Date 08/05/2020
Event Type  Injury  
Manufacturer Narrative

Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies. There have been no other events for the lot referenced. If additional information is received, it will be provided in a supplemental report upon completion of the investigation. Device not returned.

 
Event Description

It was reported that the patient's left knee was revised due to loosening of the femoral component. Surgeon reported that the distal femur metaphysis looked "disintegrated. " patient's knee construct was revised to a hinged distal femur construct. Rep confirmed there are no allegations against the revised tibial baseplate or insert. Rep provided primary and revision usage sheets, pre- and post-revision x-rays, and confirmed that no further information will be released.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameTRIATHLON CR FEM COMP #3 L-CEM
Type of DevicePROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
NA
Manufacturer Contact
alessandra chavez
2555 davie road
fort lauderdale, FL 33317
9546280700
MDR Report Key10475775
MDR Text Key205117356
Report Number0002249697-2020-01771
Device Sequence Number1
Product Code MBH
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberK141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL
Reporter Occupation
Type of Report Initial
Report Date 09/01/2020
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received09/01/2020
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? No
Device Operator HEALTH PROFESSIONAL
Device MODEL Number5510-F-301
Device Catalogue Number5510F301
Device LOT NumberJLH6D
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/05/2020
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured11/25/2019
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 09/01/2020 Patient Sequence Number: 1
-
-