MAUDE Adverse Event Report: DEPUY MITEK LLC US MINI QA+ #2/O OCORD V-5; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US MINI QA+ #2/O OCORD V-5; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE Back to Search Results
Model Number 212035
Device Problem Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/20/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint#: (b)(4).Initial reporter phone number: (b)(6).Udi: (b)(4).
 
Event Description
It was reported that during the surgery of lateral malleolar ligament repair ,the anchor¿s end tip was deformed.Another device was used to complete the surgery. there were no adverse consequences to the patient. no additional information could be provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary:according to the information provided, it was reported that during the surgery of lateral malleolar ligament repair ,the anchor¿s end tip was deformed.The complaint device was received and inspected.Visual inspection confirms that one arc of the anchor tip was bent and was detached from the shaft inserter but held in place by the suture.The complaint can be confirmed.The possible root cause can be attribute when excess torque was applied when inserting the anchor at off angle which caused the inserter tip to bend.However, it cannot be conclusively affirmed.A manufacturing record evaluation was performed for the finished device lot number:2l06440, and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINI QA+ #2/O OCORD V-5
Type of Device
SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key10481738
MDR Text Key208111276
Report Number1221934-2020-02350
Device Sequence Number1
Product Code JDR
UDI-Device Identifier10886705001835
UDI-Public10886705001835
Combination Product (y/n)N
PMA/PMN Number
K071257
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup
Report Date 08/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2022
Device Model Number212035
Device Catalogue Number212035
Device Lot Number2L06440
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2020
Date Manufacturer Received09/22/2020
Patient Sequence Number1
-
-