• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD CARELINK SMARTSYNC BASE; ANALYZER, PACEMAKER GENERATOR FUNCTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD CARELINK SMARTSYNC BASE; ANALYZER, PACEMAKER GENERATOR FUNCTION Back to Search Results
Model Number 24970A
Device Problems Capturing Problem (2891); Computer Operating System Problem (2898)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/18/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that while using the mobile programmer, the user tried to do pacing threshold test and no pacing occurred and then electrocardiogram (ecg) screen appeared to slow down to half speed.It was noted that the clinician had to shut everything down and restart and everything returned to normal.It was further noted through follow up that the patient was not pacing dependant.The programmer remains in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
The device was not returned for analysis.However, performance data collected from the device was received and analyzed.It was found that the issue appeared to be with the analyzer application and that in the process of running multiple iterations of the lead analysis amplitude threshold test, after the last test iteration, either the test encountered an error, or the analyzer was put in the background.Putting the application in the background may have caused an error, or the application encountered an error which led to led it to be shut down.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK SMARTSYNC BASE
Type of Device
ANALYZER, PACEMAKER GENERATOR FUNCTION
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
MDR Report Key10481954
MDR Text Key205282446
Report Number3004593495-2020-01000
Device Sequence Number1
Product Code DTC
Combination Product (y/n)N
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup
Report Date 10/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number24970A
Device Catalogue Number24970A
Was Device Available for Evaluation? No
Date Manufacturer Received10/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age73 YR
-
-