• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY MEMORY II DOUBLE LUMEN EXTRACTION BASKET; FFL, URETAL STONE DISLODGER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY MEMORY II DOUBLE LUMEN EXTRACTION BASKET; FFL, URETAL STONE DISLODGER Back to Search Results
Catalog Number MB-35-2X4-8
Device Problems Difficult to Fold, Unfold or Collapse (1254); Material Separation (1562)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/27/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation evaluation: our laboratory evaluation of the product said to be involved determined that the drive wire was separated from the handle.The device was returned with the basket fully retracted and a clear liquid inside of the tubing.The handle was manipulated and no movement was noted with the basket.The handle was disassembled and it was noted that the drive wire was disconnected from the handle with nesting inside the purple hub.For further evaluation of the drive wire cable and basket, the catheter was cut to push the drive wire cable out of the sheath.The basket was fully formed and intact, and evidence of solder was present on the handle cannula at the joint.No other anomalies were detected with the devices.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: the report of unable to open the basket was due to a separation of the device in the handle.The cause of the separation is unknown.The instructions for use indicates: "advance device through channel, in short increments, until basket sheath exits endoscope." the instructions for use also states: "confirm desired position of basket sheath relative to target.Advance basket out of sheath.Caution: pulling on sheath while advancing or retracting basket may damage device, rendering it inoperable." basket deployment difficulties and buckling/breaking of the drive wire can occur if the device experiences excessive pressure.Resistance in basket movement and bends in the catheter can occur if the elevator of the endoscope is used to deflect the device at a sharp angle.Prior to distribution, all memory ii double lumen extraction baskets are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During an endoscopic retrograde cholangiopancreatography (ercp) procedure, the physician used a memory ii double lumen extraction basket.The device was used for biliary stone removal in the common bile duct through the duodenum.The basket was retracted into the sheath before insertion into an endoscope and advanced to the target site.The user attempted to open the basket in the target site and it would not open.Therefore, another device was used instead to complete the procedure.This event was not reportable at this time.The device was received for evaluation on 14-aug-2020 where it was determined that the drive wire had detached from the handle [subject of report].A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEMORY II DOUBLE LUMEN EXTRACTION BASKET
Type of Device
FFL, URETAL STONE DISLODGER
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key10485519
MDR Text Key233926288
Report Number1037905-2020-00349
Device Sequence Number1
Product Code FFL
UDI-Device Identifier10827002251493
UDI-Public(01)10827002251493(17)230219(10)W4321696
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 09/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/19/2023
Device Catalogue NumberMB-35-2X4-8
Device Lot NumberW4321696
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/14/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/19/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-