Model Number 24970A
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/19/2020
Event Type  Malfunction  
Manufacturer Narrative

If information is provided in the future, a supplemental report will be issued.

 
Event Description

It was reported that the mobile programmer intermittently lost communication with an implanted device. It was noted that the mobile programmer was synced and doing wireless communication with the device and the user was on the analyzer. Communication was lost and returned intermittently and then just before testing one of the leads, it was lost completely. It was further noted that the tablet and the mobile programmer base were at the end of the table and cautery machine was on the right side of the patient. The mobile programmer remains in use. No patient complications have been reported as a result of this event.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARELINK SMARTSYNC BASE
Type of DeviceANALYZER, PACEMAKER GENERATOR FUNCTION
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY 11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY 11900
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10488943
MDR Text Key205527288
Report Number3004593495-2020-01009
Device Sequence Number1
Product Code DTC
Combination Product (Y/N)N
Reporter Country CodeUS
PMA/PMN NumberP890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE
Reporter Occupation
Type of Report Initial,Followup
Report Date 09/25/2020
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received09/03/2020
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number24970A
Device Catalogue Number24970A
Was Device Available For Evaluation? No
Is The Reporter A Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/03/2020
Was Device Evaluated By Manufacturer? No Answer Provided
Date Device Manufactured08/22/2019
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-