• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO FIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO FIBERSCOPE Back to Search Results
Model Number URF-P7
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/04/2020
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to omsc for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that as a result of multiple microbiological testing by the user facility, following microbes were detected from the sample collected from the subject device.[first time; (b)(6) 2020] unspecified microbes (7ufc/100ml) [second time;(b)(6) 2020] unspecified microbes (2ufc/100ml) [third time fourth time; (b)(6) 2020] unspecified microbes (2ufc/100ml) [fourth time; (b)(6) 2020] unspecified microbes (1ufc/100ml) the device had been manually reprocessed using peracetic acid.There was no report of infection associated with this report.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The device has not been returned to olympus medical systems corp.(omsc) but was returned to olympus france.(ofr).Ofr sent the device to a third party laboratory for microbiological testing.As a result of the testing, no microbe was detected from the sample collected from all channel of the device.The testing result cleared the french guideline.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10492459
MDR Text Key224964255
Report Number8010047-2020-06180
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 10/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P7
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-