MEDTRONIC, INC. CARELINK PROGRAMMER; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 07/20/2020
Event Type  malfunction  
Manufacturer Narrative
Product analysis: at analysis it was determined the programmer¿s stylus tip was loose and difficult to use.Analysis also noted that the power cord bay was broken, keyboard space bar was missing, keyboard hinges were broken, both rail covers were broken, hinge plate screws were missing, lower eject lever was broken on the personal computer memory card slot and the device failed left antenna test.All found defective parts were replaced and all other identified issues were resolved.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmer was returned for evaluation.There was no patient involvement.It was further reported that the programmer subsequently tested out of specification during manufacturer's analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK PROGRAMMER
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key10493043
MDR Text Key205639354
Report Number2182208-2020-01744
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090
Device Catalogue Number2090
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2020
Date Manufacturer Received08/21/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/31/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-