MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN 5000 ARCTIC SUN DEVICE Back to Search Results
Model Number 50000000
Device Problems Break (1069); Decreased Pump Speed (1500); Temperature Problem (3022)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the patient's temperature was 38c on the arctic sun device. The nurse noted the day shift nurse adjusted the target temperature from 37c to 33. 5c to get the patient's temperature down since it was not being maintained. The flow rate was 2. 0 lpm, the water temperature was 29. 8c. The event log showed an alert 113 (reduced water temperature control). Mss advised that they should not adjust target temperature if arctic sun device was not maintaining temperature. The arctic sun device needs to be taken out of service and sent to biomed labeled alert 113 (reduced water temperature control) and broken cooler. The user did not have another arctic sun device. Per follow up via nurse on (b)(6) 2020, therapy was discontinued and the arctic sun device was sent to biomed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key10494321
MDR Text Key206773583
Report Number1018233-2020-05705
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/22/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/04/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000
Device Catalogue Number50000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received12/01/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-