Medtronic received information that the tip of the pipeline flex with shield technology did not open after it was deployed, and the distal wire disconnected from the pushwire.
Re-sheathing was done more than 3 times to attempt to open and deploy it but without success.
The pipeline flex with shield technology was also noted to be shortened.
The device was removed with snare / collection equipment and replaced by a medtronic product to complete the procedure.
It was reported there were difficulties deploying in the distal region of the pipeline flex with shield technology.
The pipeline flex with shield technology was deployed less than 50%.
In an attempt to deploy the pipeline flex with shield technology, the physician pushed the device with more strength against resistance.
Ancillary devices: chikai 14 guidewire, phenom 27 microcatheter, navien 5fr guide catheter, shuttle 6fr sheath follow up response received with the following information: friction was not noticed during delivery or positioning.
The pipeline did not jump.
Multiple pipelines were not used.
The tip of the microcatheter was not under stress.
It was reported that the microcatheter dislodged naturally during deployment.
Re-sheathing was performed repeatedly.
It was confirmed that the distal wire disconnected, and the distal side was deformed.
Follow up response received with information that the catheter dislodged spontaneously (unexpectedly) during deployment.
|