DATEX-OHMEDA, INC. CARESTATION 650; ANESTHESIA GAS MACHINE Back to Search Results
Model Number 1012-9650-000
Device Problem Improper Flow or Infusion (2954)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/13/2020
Event Type  malfunction  
Manufacturer Narrative
A ge healthcare service representative performed a checkout of the system and confirmed the reported issue.The link 25 system was re-calibrated to resolve the reported issue.No report of patient involvement.(b)(4).
 
Event Description
The hospital reported the gas proportioning system was not operating as expected, causing the potential of a hypoxic mix of gas delivery.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESTATION 650
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
john szalinski
MDR Report Key10501370
MDR Text Key205917691
Report Number2112667-2020-02530
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151570
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 09/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1012-9650-000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/13/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/20/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-