Model Number 50000000
Device Problems Insufficient Cooling (1130); Decreased Pump Speed (1500); Temperature Problem (3022); Therapeutic or Diagnostic Output Failure (3023)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed. The device was not returned.
 
Event Description
It was reported that the arctic sun device was slow to cool the patient. The patient's temperature was 37c with a target temperature of 33c with an alert 114, the treatment was stopped in the event log. The flow was good. The diagnostics shows that the outlet control temperature (t1) was 37. 1c and the chiller temperature (t4) was 3. 3c. Mss advised that it seems to be a pump issue and would need to be serviced. A functional check shows that the mixing pump had failed and requested a quote for the (b)(4) service package.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key10502139
MDR Text Key207135778
Report Number1018233-2020-05740
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 04/05/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000
Device Catalogue Number50000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/17/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-