ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012453-12
Device Problem Unintended System Motion (1430)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/17/2020
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for analysis. A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event. Additionally, a review of the complaint history identified no other incidents from this lot. The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that interaction with the anatomy and/or other devices while using the kissing balloon technique resulted in the device slipping; thus resulting in the reported unintended system motion. There is no indication of a product quality issue with respect to design, manufacture or labeling.
 
Event Description
It was reported that the procedure was to treat a 70% stenosed, moderately calcified and moderately tortuos lesion in the left anterior descending diagonal artery. The 4. 0x12mm nc trek balloon dilatation catheter (bdc) slipped during kissing balloon technique on all three attempts. The procedure was successfully completed with a non-abbott same size nc balloon. There was no adverse patient effect and no clinically significant delay in the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameNC TREK CORONARY DILATATION CATHETER
Type of DeviceCORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key10502876
MDR Text Key205946184
Report Number2024168-2020-07455
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K103153
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation
Type of Report Initial
Report Date 09/08/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received09/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number1012453-12
Device Catalogue Number1012453-12
Device Lot Number90703G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/18/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/03/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-