Model Number 328512
Device Problems Leak/Splash (1354); Material Puncture/Hole (1504)
Patient Problem Needle Stick/Puncture (2462)
Event Date 08/15/2020
Event Type  malfunction  
Manufacturer Narrative
Medical device expiration date: na. A device evaluation and/or device history review is anticipated, but is not complete. Upon completion, a supplemental report will be filed. (b)(4).
 
Event Description
It was reported that syringe 0. 3ml 31ga 8mm 10bag 500 wal cannula pierced through the shield and stuck the consumer's finger. The needle hub assembly came apart on 4 occasions during use. The following information was provided by the initial reporter: material no: 328512 batch no: 0027372. It was reported that the consumer stuck her finger on a syringe with a needle thru shields. Consumer reported found 4 other syringes from this box hard to remove needle shields - when able to remove these syringe needle hub assembly removed with the shields.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYRINGE 0.3ML 31GA 8MM 10BAG 500 WAL
Type of DevicePISTON SYRINGE
Manufacturer (Section D)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer (Section G)
BD MEDICAL - DIABETES CARE
1329 west highway 6
holdrege NE 68949
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652341
MDR Report Key10505618
MDR Text Key207156995
Report Number1920898-2020-01170
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K024112
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 09/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number328512
Device Catalogue Number328512
Device Lot Number0027372
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/02/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/19/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/14/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 09/08/2020 Patient Sequence Number: 1
-
-