Model Number RT265
Device Problem Gas/Air Leak (2946)
Patient Problem No Patient Involvement (2645)
Event Date 08/10/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint rt265 infant dual-heated evaqua2 breathing circuit was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is based on the information provided by the customer and our knowledge of our product.Results: the customer reported that the rt265 circuit failed the ventilator leak test.Conclusion: without the complaint device, we are unable to determine the cause of the reported event.All rt265 infant dual-heated evaqua2 breathing circuits are visually inspected and pressure and flow tested during production, and those that fail are rejected.The subject infant breathing circuit would have met the required specifications at the time of production.Our user instructions that accompany the rt265 infant dual-heated evaqua2 breathing circuit state the following: "check all connections are tight before use." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." "set appropriate ventilator alarms.".
 
Event Description
A distributor reported on behalf of a healthcare facility in (b)(4), via a fisher & paykel healthcare field representative that a rt265 infant dual-heated evaqua2 breathing circuit failed the leak test before use.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT DUAL HEATED EVAQUA2 BREATHING CIRCUIT
Type of Device
BZE
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
MDR Report Key10505920
MDR Text Key206194709
Report Number9611451-2020-00766
Device Sequence Number1
Product Code BZE
UDI-Device Identifier09420012430205
UDI-Public010942001243020510210101708411200221
Combination Product (y/n)N
PMA/PMN Number
K103767
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial
Report Date 08/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRT265
Device Catalogue NumberRT265
Device Lot Number2101017084
Was Device Available for Evaluation? No
Date Manufacturer Received08/11/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/21/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-