BLUE BELT TECHNOLOGIES REAL INTELLIGENCE ROBOTIC DRILL ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB10013
Device Problem Device Dislodged or Dislocated (2923)
Patient Problems Failure of Implant (1924); No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/12/2020
Event Type  Injury  
Event Description
It was reported that during a cori tka procedure, the first robotic drill could not seat the bur in the carriage. They swapped to another robotic drill that did not seem to mill as it should, even though they did not receive any errors. It did not appear to be cutting bone efficiently. The bone was very hard and the surgeon moved between making saw cuts and cleaning up cuts with bur, but it did not appear to cut much. There was a change to manual procedure and a delay greater than 30 minutes. No other complications have been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameREAL INTELLIGENCE ROBOTIC DRILL
Type of DeviceORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
rick confer
2828 liberty avenue
suite 100, PA 15222
MDR Report Key10514169
MDR Text Key206312482
Report Number3010266064-2020-01755
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K201022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Type of Report Initial,Followup
Report Date 10/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/10/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Model NumberROB10013
Device Catalogue NumberROB10013
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/03/2020
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/27/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 09/10/2020 Patient Sequence Number: 1
-
-