Manufacturing record was reviewed, there were no nonconformance related to this lot, therefore, supporting the device met material, assembly and performance specifications.A follow-up report will be issued after the investigation is complete.
|
Device returned for investigation.The device lot history record review indicated no non-conformities related to this lot, therefore, supporting the device met material, assembly, and performance specifications prior to shipment.The event description states that the tip broke off before being inserted into the patient.The account stated that it is unknown if the catheter was broken in the package or after it was removed from the package.Upon inspection of evaluation device, no blood particulates were found on the catheter.The weld joint break was confirmed.The tip of the catheter was intact, and no damages were found.The likely cause of weld joint break could be use related during preparation and handling.Per ifu, a warning and precaution are stated as of the following: exercise care in handling the catheter during a procedure to reduce the possibility of accidental breakage or kinking.Do not apply torque to the catheter during delivery, as catheter damage may result.Based on the information and returned product evaluation, the most likely root cause of the issue is unintended use error.
|