Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed silicone damage at the magnet pocket, top and bottom covers, and near the array.In addition, the electrode ground ring was migrated prior to receipt.These are believed to have occurred during revision surgery.The photographic imaging inspection revealed broken electrode wires at the fantail and near the electrode ground ring.These are believed to have occurred during revision surgery.System lock was verified.The device passed the electrical test performed.The residual gas analysis could not be performed due to damage during failure analysis.This device was explanted for medical reasons.The device passed the electrical tests performed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|