• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS BROVIAC CV CATHETER, SINGLE-LUMEN, 2.7F; CHRONIC CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS BROVIAC CV CATHETER, SINGLE-LUMEN, 2.7F; CHRONIC CATHETER Back to Search Results
Model Number 0600040
Device Problems Break (1069); Fluid/Blood Leak (1250); Fracture (1260)
Patient Problem Edema (1820)
Event Date 07/10/2020
Event Type  Injury  
Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 02/2022).
 
Event Description
It was reported that post chronic catheter implantation, the device allegedly broken.It was further reported that patient developed edema under the skin and infiltration of the skin.The current patient status is unknown.
 
Event Description
It was reported that approximately one week post chronic catheter implantation, the device allegedly broken.It was further reported that patient developed edema under the skin and infiltration of the skin.The device was removed and replaced.The current patient status is unknown.
 
Manufacturer Narrative
H10: manufacturing review: a lot history review was performed.This is the only complaint to date for this lot number.Therefore, a device history record review is not required.Investigation summary: the investigation is inconclusive for the reported break issue, as the device was not returned for evaluation.A definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 02/2022),g4.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROVIAC CV CATHETER, SINGLE-LUMEN, 2.7F
Type of Device
CHRONIC CATHETER
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key10520179
MDR Text Key206486974
Report Number3006260740-2020-03227
Device Sequence Number1
Product Code LJS
UDI-Device Identifier00801741036309
UDI-Public(01)00801741036309
Combination Product (y/n)N
PMA/PMN Number
K830256
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/11/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0600040
Device Catalogue Number0600040
Device Lot NumberREBP0963
Was Device Available for Evaluation? No
Date Manufacturer Received11/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-