Catalog Number H938740
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Patient Involvement (2645)
Event Date 08/05/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a 2000ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag was leaking from the front side of the bag, through the printed volumetric scale.This occurred during setup prior to patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H4: device manufactured between january 24, 2020 to january 25, 2020.H10: the device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A functional testing was performed which revealed a leak in the front side of the bag approximately at the 1600 ml fixed print scale area.Upon magnified inspection verified a small tear/hole in front of the bag where the leak occurred.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2000 ML TPN
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10524557
MDR Text Key206732147
Report Number1416980-2020-05634
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477244
UDI-Public(01)00085412477244
Combination Product (y/n)Y
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date12/31/2022
Device Catalogue NumberH938740
Device Lot Number60222962
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2020
Date Manufacturer Received10/06/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-