The device was received for evaluation.
During visual inspection, both buttresses were observed to still be contained within the sheath with the paper backer still intact.
The anvil side tissue buttress was partially displaced from under the sheath rails on the distal side of the sheath.
The paper backer was removed from both the cart and anvil sides with no further disturbance to the tissue position, loaded onto a stapler, and fired with no adhesion issues.
The reported condition was verified.
The cause of the condition was due to user error.
Per the labeling for peri-strips, it states, the paper backer to be removed prior to use.
When loading the buttress onto the stapler, the customer likely grabbed both the tissue and tab of the paper backer and pulled the tissue out of the sheath.
If this issue occurs, the customer may reposition the buttress once it¿s on the stapler, similar to what was done during sample evaluation, to ensure staple line coverage (step 18).
Should additional relevant information become available, a supplemental report will be submitted.
|
During device evaluation, it was found the user failed to follow instructions during use of the peri-strips.
When loading the buttress onto the stapler, the customer likely grabbed both the tissue and tab of the paper backer and pulled the tissue out of the sheath.
If this issue occurs, the customer may reposition the buttress once it's on the stapler, similar to what was done during sample evaluation, to ensure staple line coverage.
The event occurred prior to patient use; therefore, there was no patient involvement.
No additional information is available.
|