SMITH & NEPHEW, INC. REFLECTION REAMER DOME 57MM; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
Model Number 71362757 |
Device Problem
Material Separation (1562)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 08/14/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that, during thr procedure, the "+" where the reamer dome 57 mm connects to the reamer shaft broke off.The procedure was finished using a smith and nephew backup device.Surgery was not delayed.The patient was not harmed beyond the described event.
|
|
Manufacturer Narrative
|
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The reamer shaft had separated from the reamer.The device was manufactured in 2019 and shows signs of extensive use.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|